ANNEX VI

CONTAINMENT FOR INDUSTRIAL PROCESSES

(Article 4 (1) and Article 16 (2) (a))

Group 1 biological agents

For work with group 1 biological agents including life attenuated vaccines, the principles of good occupational safety and hygiene should be observed.

Group 2, 3 and 4 biological agents

It may be appropriate to select and combine containment requirements from different categories below on the basis of a risk assessment related to any particular process or part of a process.

Containment measures

Containment levels
2 3 4

1. Viable organisms should be handled in a system which physically separates the process from the environment

Yes Yes Yes

2. Exhaust gases from the closed system should be treated so as to:

Minimize release Prevent release Prevent release

3. Sample collection, addition of materials to a closed system and transfer of viable organisms to another closed system, should be performed so as to:

Minimize release Prevent release Prevent release

4. Bulk culture fluids should not be removed from the closed system unless the viable organisms have been:

Inactivated by validated means Inactivated by validated chemical or physical means Inactivated by validated chemical or physical means

5. Seals should be designed so as to:

Minimize release Prevent release Prevent release

6. Closed systems should be located within a controlled area

Optional Optional Yes, and purpose- built

(a) Biohazard signs should be posted

Optional Yes Yes

(b) Access should be restricted to nominated personnel only

Optional Yes Yes, via airlock

(c) Personnel should wear protective clothing

Yes, work clothing Yes A complete change

(d) Decontamination and washing facilities should be provided for personnel

Yes Yes Yes

(e) Personnel should shower before leaving the controlled area

No Optional Yes

(f) Effluent from sinks and showers should be collected and inactivated before release

No Optional Yes

(g) The controlled area should be adequately ventilated to minimize air contamination

Optional Optional Yes

(h) The controlled area should be maintained at air pressure negative to atmosphere

No Optional Yes

(i) Input air and extact air to the controlled area should be HEPA filtered

No Optional Yes

(j) The controlled area should be designed to contain spillage of the entire contents of the closed system

No Optional Yes

(k) The controlled area should be sealable to permit fumigation

No Optional Yes

(l). Effluent treatment before final discharge

Inactivated by validated means Inactivated by validated chemical or physical means Inactivated by validated chemical or physical means